New TopShim run:

version = topshim 2.0.0
date = 2022-04-06
time = 14:17:59.904
command = topshim 

FINE SHIMMING

Parameters:

probe = Z108618_0165
solvent = DMSO
shim nucleus = 2H
nucleus optimised for = 1H
o1p (from lock) = 2.49 ppm

Results:


1D Shimming:
Iteration  1: FWHM Before= 0.7 Hz, Target= 0.4 Hz, After= 0.3 Hz
Iteration  2: FWHM Before= 0.3 Hz, Target= 0.3 Hz, After= 0.6 Hz
Iteration  3: FWHM Before= 0.6 Hz, Target= 0.4 Hz, After= 0.4 Hz
Iteration  4: FWHM Before= 0.4 Hz, Target= 0.4 Hz, After= 0.5 Hz
Iteration  5: FWHM Before= 0.5 Hz, Target= 0.4 Hz, After= 0.5 Hz
Iteration  6: FWHM Before= 0.5 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  7: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Slow convergence detected. Shimming was aborted.
Using the best result as the final shim state.

Lineshape quality factor (1D spectrum, spinning) = 0.506

shim changes: 
Z	+9
Z2	+109
Z3	-410
Z4	-1008
Z5	+2168
Z6	+5954

duration = 1 min 6 sec
completed successfully
finished Wed Apr  6 14:19:05 2022


========================================================================